Volume 14, Issue 8 , Pages 789-797, August 2001
Radiofrequency spectral analysis of attenuated ultrasound signals in experiments with echo contrast microbubbles☆
Abstract
Conventional gray-scale myocardial contrast echocardiography cannot distinguish perfused but attenuated from nonperfused myocardium because both may appear similar at low image intensity. We hypothesized that with radiofrequency spectral analysis of attenuated ultrasound signals, the harmonic-to-fundamental frequency ratio of the peak power spectrum (HFRP) could determine the presence of contrast microbubbles. We measured frequency responses of Optison microbubbles at defined degrees of ultrasound signal attenuation with different formulations of silicone (55D, 80A, and 3M); gray-scale intensities of Optison plus water compared with degassed water were analyzed at different attenuation settings (−25, −32, and −44 dB, respectively). HFRP values of Optison plus water were significantly higher than reference values of degassed water at each attenuation setting (55D, −14 ± 2 dB versus −30 ± 2 dB, P < .001; 80A, −19 ± 2 dB versus −30 ± 3 dB, P < .01; 3M, −22 ± 2 dB versus −30 ± 3 dB, P < .05), even though conventional videodensitometric analysis could not distinguish them. HFRP analysis objectively detects microbubbles in clinically relevant conditions of attenuation. (J Am Soc Echocardiogr 2001;14:789-97.)
To access this article, please choose from the options below
☆ Nothing in this publication implies that Mayo Foundation endorses any products of Mallinckrodt Medical.
PII: S0894-7317(01)17323-2
© 2001 American Society of Echocardiography. Published by Elsevier Inc. All rights reserved.