Volume 24, Issue 3 , Pages 350.e5-350.e6, March 2011
Acute Aortic Regurgitation Secondary to Disk Embolization of a Björk-Shiley Prosthetic Aortic Valve
Having passed the 30th anniversary of the first implantation of a Björk-Shiley convexo-concave tilting mechanical valve, recognition of the life-threatening complication of strut fracture is not widespread. The authors report the case of a 48-year-old man with acute-onset chest pain and dyspnea found to have strut fracture and disk embolization of a 26-year-old Björk-Shiley prosthetic aortic valve. The value of echocardiography in the diagnosis of this condition is discussed.
Keywords: Björk-Shiley, Aortic insufficiency, Strut fracture
To access this article, please choose from the options below
PII: S0894-7317(10)00554-7
doi:10.1016/j.echo.2010.07.001
© 2010 American Society of Echocardiography. Published by Elsevier Inc. All rights reserved.